Relative humidity measurement |
Measurement rannge |
0 ... 100 % RH
|
Accuracy at +20 °C (68 °F) |
± 1.0 % RH (0 ... 90 % RH) |
Accuracy at +20 °C (68 °F) |
± 1.7 % RH (90 ... 100 % RH) |
Sensor (general use) |
HUMICAP® 180 and HUMICAP® 180R |
Sensor (demanding chemical environment) |
HUMICAP® 180L2 |
Temperature measurement |
Measurement range
HMT361
HMT363
HMT364/365/367/368
|
-40 ... +60 °C (-40 ... +140 °F)
-40 ... +120 °C (-40 ... +248 °F)
-40 ... +180 °C (-40 ... +356 °F)
|
Typical accuracy of electronics at +20 °C (68 °F) |
± 0.2 °C (± 0.36 °F) |
Sensor |
Pt 1000 RTD 1/3 Class B IEC 751 |
Other optional variables available |
Dewpoint temperature, mixing ratio, absolute humidity, wet bulb temperature. |
Operating environment |
Temperature
for electronics
with display
storage |
-40 ...+ 60 °C (-40 ... +140 °F)
-20 ...+ 60 °C (-4 ... +140 °F)
-40 ...+ 70 °C (-40 ... +158 °F)
|
Inputs and outputs |
Operating voltage
with serial port (servicemode)
|
12 ... 28 V
15 ... 28 V
|
Analog outputs |
two-wire 4 ... 20 mA, one standard, one optional |
Digital outputs |
RS-232C serial output for service use |
Mechanics |
Weight (housing)
|
950 g
|
Material (housing) |
G-AlSi 10 Mg (DIN 1725) |
Housing classification |
IP66 (NEMA 4X) |
Options and accessories |
Additional analog output
|
4 ... 20 mA
|
Duct mounting installation kit (for HMP363/367) |
210697 |
Installation flange (for HMP365) |
210696 |
Ball valve set (for HMP368) |
BALLVALVE-1 |
Serial interface cable for PC |
25905ZZ |
Protection cover |
214101 (for use in the presence of combustible dust) |
Galvanic isolator |
212483 |
Zener barrier (USA & Canada) |
210664 |