Stable and Accurate Precision
Thermistors |
- Highly Accurate
temperature measurement
- Ultra stable
- Custom packaging
- Available in bulk
packs of 100 each
|
|
Precision
Interchangeable Epoxy Thermistors YSI thermistors provide highly accurate
and stable temperature sensing for applications of temperature measurement,
control, indication and compensation. typical uses include precise
measurements without the necessity of individual circuit calibration
and with the advantage of precision interchangeability of sensors.
Precise cold junction compensation of thermocouples may be designed
directly without "bread boarding" after mathematically deriving the
circuit because of the superior interchangeability of YSI precision
thermistors. Two interchangeability tolerances and two thermistor
configurations are offered. Teflon encased thermistors allow exposure
to hostile environment such as conductive or corrosive liquids and
particulate suspensions. A stiff wire placed in the tube also allows
the thermistor leads to be formed to various shapes with slight finger
pressure.
Time Constant: The time required for a thermistor to indicate
63% of a newly impressed temperature is the time constant. For a thermistor
suspended by its leads in a well-stirred oil bath it is 1 sec. max.
for standard thermistor and 2.5 sec. max. for teflon encased thermistors.
In still air it is 10 sec. max. for standard thermistors and 25 sec.
max. for Teflon encased thermistors.
Dissipation Constant: The power
in mW required to raise a thermistor 1°C above surrounding temperature
is the dissipation constant. For all thermistors suspended by their
leads in a well stirred oil bath it is 8mW/°C min. or 1mW/°C in still
air.
Interchangeability: See Table = +/- 0.1°C Tolerance from 0 to
70°C = +/- 0.2°C Tolerance from 0 to 70°C
Stability: YSI thermistors
are chemically stable and not significantly affected by aging or exposure
to strong nuclear radiation.
Operating: Typical Thermoelectric Drift Temperature 10 Months
100 Months 25°C <0.01°C <0.01°C 70°C <0.12°C <0.01°C 100°C <0.15°C
<0.03°C 150°C <0.20°C <0.08°C 200°C <0.20°C <0.60°C |
|
|
|
|
|
|
|
Important
Notice: By purchasing products from this Site, you acknowledge and
agree that you have read and accept the Terms and Conditions of this
online store. For more information email us at: info@iprocessmart.com |
|
|
|
|
|
|
|
|
|
|
|
|
|
Technical
help and information, Conversions acd corrosion charts, Standards,
heater and motor sizing, Reference iIndustrial iInformation, |
|
|