• Self-contained, ergonomic,
easy to operate
• Fast, easy
connection to computer or printer
• Selectable
sample time, count data, programmable delay
• Set sample
times to cubic feet or liters
• Data displayed
in totalize or concentration modes
• Built-in, programmable
alarms
• Built-in purge
filter protects the cleanroom from pump exhaust
• Automatic turn--off
when battery voltage drops below safe operating level
• Meets the most
stringent performance standards set by the industry
Specifications
6 Size Channels: Standard:
0.3,0.5,0.7,1.0,2.0,5.0 µm
IAQ: 0.5,0.7,1.0,2.0,5.0,10.0
µm
Custom: 0.3 to
20 µm available
Flow Rate: 0.1
cfm (2.83 L/min)
Light Source:
Laser diode; index guided (26,000 MTBF at 25 °C)
Calibration:
PSL particles in air (NIST traceable)
Counting Efficiency:
50% @ 0.3 µm; 100% for particles >0.45 µm (per JIS B9921:1997)
Zero Count: 1
count/5 minute (per JIS B9921:1997)
Coincidence Loss:
5%at 2,000,000 particles per ft3
Relative Humidity:
± 7%,20%to 90%,non-condensing
Temperature:
± 3 °C,10 °C to 40 °C (50 °F to 104 °F)
Data Storage:
500 sample records (rotating buffer)
Data Recorded:
Date, time, counts, relative humidity, temperature, sample volumes,
alarms, label
Display: Graphic
liquid crystal with back light
Alarms: Counts,
low battery, sensor fail
Count Modes:
Concentration, totalize, audio
Delay Time: 0
to 24 hours
Sample Inlet:
Isokinetic probe
Interface: RS-232
and RS-485 via RJ-45
Vacuum Source:
Internal pump flow controlled
Housing: Injection
molded plastic
Dimensions: 4.5"w
x 8.25"l x 2.25"d (11.43 x 20.96 x 5.72 cm)
Weight: 2.2 pounds
(1.0 kg)
Environmental:
Operating: 10 °C to 40 °C (50 °F to 104 °F), 20% to 90% relative
humidity, non-condensing
Storage: -10
°C to 50 °C (14 °F to 122 °F), Up to 90% relative humidity, non-condensing
Power: AC Adapter,
12 VDC at 2.5 A, 90 to 250 VAC, 50 to 60 Hz
Rechargeable
Battery: NiMH,4.8 V at 4.5 Ah; replaceable
Continuous Operating
Time: 8 hours
Recharge Time:
2 hours
Standards: CE, JIS
B9921:1997
Accessories Included:
Certificate of Calibration (NIST); Windows compatible software download
utility;DB9 to RS-232 adapter and cable; Isokinetic Probe; High
Purity Tubing; 1/8 in. Hose Barb Adapter; Power Supply; Operation
Manual
Optional Accessories
Zero Count Filter
Carrying Case
(soft-sided or hard molded plastic) data
Portable serial
printer w/cable, 2 rolls paper
Option "EX
" – 2,000 records
Optional particle
size range – up to 20 µm
|