An
Inexact Science
OPC manufacturers generally specify
sensitivity and counting efficiency on the basis of ideal
test particles that are transparent and spherical. Most
often, polystyrene latex (PSL) spheres, with a refractive
index of about 1.59, are used for testing. In the particle
counting industry there is a tendency to emphasize PSL sphere
sensitivity and skirt the issue of OPC sizing accuracy and
sensitivity with particles found in the real world. Unfortunately,
real world particles come in a wide variety of shapes and
refractive indices, leading to a significant degradation
of sensitivity, resolution and accuracy. Sizing real world
particles is an inexact science.
Particle
Diameter
Resolution - the smallest detectable
particle size difference. It is the ratio of the standard
deviation (a) to the diameter (D) expressed as a percentage:
% Res = s
x 100%
D
Resolution
is a function of the width of the bellshaped curve. It is
also referred to as "coefficient of variation, relative
precision and relative standard deviation".
Precision
- the standard deviation (d) of repeated measurements of
the same size monodispersed spheres:
where: D1
= the i-th measurement of particle diameter
(arithmetic mean of N measures)
N = total number of measurements
Reproducibility
(also called repeatability and calibration stability) -
the extent to which an OPC will give the same sizing and
counting response to the same diameter PSL spheres over
a long period of use.
Sensitivity
- the smallest size particle an OPC can detect at a specified
counting efficiency, e.g., 0.3 micron at 50% counting efficiency.
Counting
efficiency - the detected particle concentration divided
by the true concentration (as measured by a hypothetically
perfect instrument). This curve provides useful information
regarding the sensitivity and resolution of the instrument.
False count
rate - the counts per unit volume using perfectly filtered
air at a specified flowrate.
Signal -
the magnitude of the sensed scattered light produced only
by the passing of a particle through the view volume. Size
Is deduced from the signal magnitude. Noise is the opposite
of signal in that it is produced by anything but a particle
in the view volume. A high signal-to- ratio implies low
false count rate.
Counting
Efficiency, Sensitivity and Resolution
As an aid to arriving at a definition
of counting efficiency, let us assume the presence of an
ideal reference particle counter. See block diagram on next
page. This counter can "see" every particle passing
through the view volume to a diameter much lower than the
lowest detection limit of the UPC under test. Typically,
this instrument is a condensation nucleus counter (CNQ or
an OPC with a verified counting efficiency of 100% at the
lower detection limit of the OPC under test. However, a
CNC only counts particles above a given size corresponding
to a preset threshold (e.g. 0.01 micron); it cannot size
particles. A reference CNC must be used with an electrostatic
classifier to analyze particles by controlled deflection
in an electrostatic field.
An aerosol
carrying monodispersed PSL spheres is generated by the atomizer.
The aerosol is mixed with filtered air in the mixing chamber.
The OPC under test and the reference counter simultaneously
sample the monodispersed spheres at the same concentration.
As smaller and smaller monodispersed spheres are introduced,
there is a point where the OPC under test fails to detect
all the particles that the reference instrument is sensing.
Further reduction in particle size results in the eventual
loss of particle detection.
Counting
efficiency is expressed as follows:
CE = Nm/N,
x 100%
where Nm
is the measured concentration and N1 is the true
concentration as measured by the reference instrument.
|
|
|
Setup
for Measuring Counting Efficiency
In all cases, the threshold
of the counter's detector is set to sense those monodisperse
particles which fall in the upper half of the bell
curve (those particles to the right of the intersection
point of the three curves shown below). Smaller particles
in the lower half of the curve (those to the left
of the intersection point) are intentionally not sensed
or counted.
Curve
A represents the hypothetical case where the PSL spheres
am ideal (exactly the same size with no dispersion)
and the particles are sized perfectly by the OPC (also
with no dispersion). Here, the 0, 50 and 100% efficiency
points lie on the same vertical line. In this case
the bell shaped curve is simply a straight line. Unfortunately,
particle counters do not exhibit such steep function
efficiency curves.
Curve
B, exhibiting a relatively steep slope, is typical
of a counter with good resolution, whereas curve C
is representa- of a counter with poor resolution.
Size dispersion is much smaller in the case of the
superior instrument (see curves in figure inset below).
Let's
examine the elements that determine the slope in practice.
The largest contributor to poor resolution is the
lack of uniformity of light intensity across the view
volume. With any optical system, it is difficult to
collimate a light beam down to a small area to achieve
good sensitivity and, at the same time, maintain perfectly
uniform intensity across this area. Any non-uniformity
causes a discrepancy between sizing a particle passing
right through the middle of the view volume and one
that passes through one edge of the volume (being
mistaken for a smaller particle). Variations in flow
rate also contribute to wider dispersions resulting
in degraded resolution. Other contributors, such as
photodetector and amplifier stabilities are usually
negligible in the typical particle counter.
If
the counter thresholds corresponding to curves B and
C had been set to anything but 50%, the particle counts
between the two instruments would be in total disagreement.
Only at 50% counting efficiency would two instruments
with different resolutions count exactly half the
monodispersed particles introduced.
If,
for example, the thresholds were set for 90% counting
efficiency at 0.12 microns as shown in the figure
below, the curve C counter would outcount the curve
B counter by a wide margin.
A quick
and easy way to assess resolution is to compare the
50 and 100% points. In a typical Met ONE 0.1 micron
counter this difference is about 0.015 microns, corresponding
to a resolution of about 5%. This represents good
resolution in the OPC industry. A more effective approach
is to obtain (or generate) the counting efficiency
curve for the instrument in question and determine
the size spread between the 10 and 90% points. This
provides a solid basis for specifying and comparing
OPCS. In a typical cleanroom, the distribution of
particles vs particle size follows a power law function
as shown in the next curve. Given this distribution,
theory predicts that a poor resolution instrument
will actually out-count one with good resolution at
sizes below the threshold settings. Multichannel counters
have a number of threshold settings, each set for
50% counting efficiency at the designated size. Note
in the following set of curves the change in slope
(and the apparent degradation of resolution) as particle
size increases. This is due to the change in power-law
exponent of the intensity of scattered light versus
particle size.
|
|
|
|
In
setting the threshold at the lower sensitivity
level, the signal level must be above the noise
level. If. in order to stretch sensitivity,
the threshold is set too close to the noise
level, the false count rate (zero particles)
will increase. To avoid this, always make sure
that sensitivity and counting efficiency are
accompanied by a minimum acceptable false count
specification.
Particles
in the Real World
Measurements with
ideal (PSL) spheres provide us with a powerful
tool for assessing the sensitivity, accuracy,
resolution and false count level of a counter.
This calibration technique serves two purposes:
1) Gives comparative evaluations of a wide variety
of counters, 2) provides a measure of how. well
a counter maintains its calibration (reproducibility).
Parameters
such as accuracy, counting efficiency, and resolution
are very important in the process of PSL-based
measurements. For this reason a somewhat detailed
study was given in the "Definition of Terms"
section. However these parameters become meaningless
when it comes to measuring real world particles.
Particles found in the cleanroom have a wide
variety of shapes and refractive indices. This
leads to a significant degradation of sensitivity,
accuracy and resolution.
In
an article in the periodical, Microcontamination,
(ref. no. 8) Stuart Hoenig makes the point that
"At 0. 1 microns, optical scattering....is
strongly dependent on the particle's shape,
color and electromagnetic characteristics"
which reinforces the fact that significant error
exists in sizing real world particles.
|
|
|
|
To
be able to "see" all of the particles in the view
volume, the volume must be totally illuminated. Such a sensor
is called "volumetric"; one that is in line with
the flow and sees a sample of the aerosol is termed "in-situ".
Some of the most sensitive counters available detect, for
example, only 80% of the particles introduced and die curves
am "normalized" to the 100% point when constructing
the counting efficiency curves. All counting efficiency
curves published by MET ONE reflect the true count This
fact must be considered when ordering a counter.
Some elements
commonly found in contaminants, the by-products of processes
like ionic etching and vacuum deposition, are:
aluminum copper nickel
boron fluoride phosphorus
calcium gold potassium
carbon iron sodium
chlorine lithium sulfur
chromium manganese tin
These elements
exhibit a wide range of refractive indices. Some am highly
reflective while others absorb most of the incident light
energy.
Particle
sizing errors due to changes in refractive index encountered
with particles from the elements listed above are shown
in the curves below. This is a family of response curves
for a 90* (off-axis), 24* collection angle, aerosol counter
for spherical particles of different diameters (0.1 to 10
µ) and refractive indices (in = 1.2 to 2.4). These curves,
derived from reference no. 11. are for nonabsorbing particles
only (except carbon). As indicated, a 0.2 µ particle at
in = 1.2, scatters about the same amount of light to the
detector as a 0.1 µ particle at m = 2.4. This represents
an accuracy of about 215% according to the definition of
terms presented earlier. This number could change even more
dramatically if the effects of absorption and differences
in particle shape am considered.
In the case
of carbon particles, which are highly absorptive, the deviation
in respect to the in = 2.4 curve increases rapidly for particles
above 0.2 µ, with the error approaching 250% above I µ (see
curves). 'Me foregoing verifies our earlier statement regarding
the inexactness of sizing real world particles with an optical
particle counter.
For practical
purposes, OPCs are calibrated with ideal particles having
a refractive index between 1.5 and 1.6. The size measured
by the OPC is then an "equivalent PSL diameter"
or an "equivalent DOP diameter", depending on
the calibrating aerosol used.
The magnitude
of error in sizing real world particles with an OPC would
appear to be a discouragement in attempting to set up an
effective cleanroom quality assurance program. Actually
the outlook is not as bleak as might be expected at first
glance. It turns out that the optical particle counter can
function as a surprisingly effective tool in the cleanroom
if used in a protocol which has been evolved by cleanroom
professionals over the years.
How
To Use An OPC In A Cleanroom
The OPC has two basic functions
in the cleanroom. The first is to certify the cleanroom
to meet standards established by FED-STD-209D. The second
function is to support a quality maintenance program in
the cleanroom.
Cleanroom
Certification
Federal Standard 209D establishes
classes of air cleanliness for airborne particulate levels
in cleanrooms. It also prescribes methods for class verification
and monitoring of air cleanliness.
The class
number is based on the number of 0.5 micron particles per
cubic foot (refer to table below). A Class 100 cleanroom,
for example, must have no more than 100 0.5 micron particles
per cubic foot in order to meet the requirements for certification.
Measurements can also be taken on any of the remaining size
ranges from 0.1 to 5.0 microns to establish the cleanroom
class.
The first step is to calibrate the particle counter with
PSL spheres according to the guidelines listed in Appendix
B of FED-STD-209D.
Then, with
a particle counter, measure particle concentration at various
points in the cleanroom for compliance with Section 5 of
FED-STD-209D. Number of locations, samples, the sample time
and minimum volume per sample are also specified in Section
5.
Perform a
statistical analysis in conformance with Section 5.4 of
the Federal Standard. First, the average particle concentration
for each location is calculated. Then the mean of the averages
for all locations is computed. Finally, the standard deviation,
the standard error and the 95% upper confidence level (UCL)
is calculated. If the UCL is found to be not more than the
particle limit shown in the table, the cleanroom meets the
requirements for certification.
A Federal
Standard 209D Calculations option is available for Met One
counters that will automatically calculate all the statistical
data required for compliance with FED-STD-209D.
A revised
version of FED-STP-209D is about to be issued. The revisions
may have impact on your operation. Contact Met One for an
up-to-date status on this standard.
Air
Quality Maintenance
Monitoring air quality with the
particle counter to support an air quality maintenance program
is more involved than cleanroom certification. The goal
is to eliminate "killer defect" particles which
can destroy product yield. Particles whose size is about
one-tenth (or larger) of the minimum line width on a semiconductor
wafer fall into this category.
Here, the
user must be more aware of the magnitude of particle sizing
errors in the real world, due to variations in shape and
refractive index. Instead of trying to size particles precisely,
establish particle concentration reference levels and correlate
these levels with product yields. At this point, the exact
sensitivity (whether it was 0.1 or 0.15 µ), as measured
earlier with PSL spheres, becomes insignificant. What is
important is the ability of the counter to hold its calibration
over the long term.
During a
particle shedding event, particles are generated in a wide
distribution of sizes. If there are 0.1 micron particles
in the sample, you can be sure there will also be for 02
micron particles present. Considering this factor, the second
decimal place in measuring micron sensitivity with PSL spheres
should not be overemphasized. Calibration stability is more
important.
Step I in
a typical cleanroom monitoring procedure is to- establish
a reliable zero count (false count level) using a quality
filter on the counter's aerosol inlet port. Repeated checks
that show -consistently low counts will give you confidence
that you have .attained an acceptable level.
Step 2 is
to establish a particle concentration .baseline for each
station to be monitored in the cleanroom. Even though the
exact size of particles counted is unknown because of differences
in shape and refractive indices, a reference level can be
created. Familiarize yourself with the levels associated
with the various processes and determine empirically what
level is acceptable and what levels begin to reduce the
yield.
Step 3 is
to recalibrate with PSL spheres from time to time until
you acquire confidence in the counter's ability to hold
its calibration in the working environment.
Summarizing,
instead of an absolute particle sizer and counter, the OPC
can be used more effectively as an early warning trend indicator
or burst detector. This will allow you to shut down a process
if the concentration level exceeds a preset threshold. Thus
the OPC can function more as a process tool than an environmental
tool in your cleanroom quality assurance program.
Stability
Considerations
Stability in the working environment
must be considered at least as important as sensitivity.
It is extremely important that the counter maintain its
calibration over the long term, otherwise, particle concentration
baselines become meaningless.
Typically,
the wavelength of light used in OPCs ranges from about 0.63
to 0.83 micron. As particles become smaller than a wavelength,
the amount of light they scatter into the detector collection
optics drops off rapidly. Referring to the Particle Size
vs. Energy Curve illustrated earlier, the detected light
energy falls off exponentially with decreasing particle
size. At 0.3 micron the detected energy drops off as a function
of about the 4th power of particle size; at 0. 1 micron
the detected light energy drops off as a function of about
the 5th to 6th power of particle size.
For example,
to upgrade the sensitivity of a counter from 0.2 to 0.1
micron requires about a 17-fold increase in light power
focused into the view volume. To achieve a sensitivity approaching
0.1 micron requires a well-designed laser/optical system
with a narrow optical bandwidth (sometimes referred to as
"high Q") in order to develop high light intensity
in the view volume. Making the optical bandwidth too narrow
Q too high) in order to achieve high sensitivity can actually
lead to calibration instability (in the presence of mild
shock or vibration) with the attendant loss of sensitivity.
Each manufacturer
is faced with making a trade-off between high sensitivity
and reproducibility in the working environment. At MET ONE
we put the emphasis on stability and shock-resistant reproducibility
in a tough environment at the expense of a slight sacrifice
in sensitivity.
If a manufacturer
decides to make the trade-off favoring sensitivity at the
expense of calibration stability, the result is a sensitive,
"highly tuned" laboratory instrument which can
easily slip out of calibration with normal use. Simply moving
such an instrument from one bench to another can degrade
its detection limit performance.
In this industry
there is a tendency to tout OPC performance solely in terms
of PSL sphere micron sensitivity at 50% counting efficiency
(e.g. 0.10 micron @ 50% counting efficiency). In the world
of particle counting in a working environment, sensitivity
alone is not a meaningful number. It does not give you enough
information to determine if you are dealing with a highly
tuned laboratory instrument or a durable and reliable workhorse.
Considering
the degradation in particle counter sensitivity when dealing
with real world particles, MET ONE does not put undue emphasis
on the second decimal place of the particle size sensitivity
specification. Our position is very clear: a counter that
maintains a long-term sensitivity of, for example, 0.12
microns (50% C.E.) in a rough environment is far superior
to an instrument that begins with a sensitivity of 0.10
microns (50% C.E.) and slips out of calibration the first
time it is moved to a new location.
Summary
1) Specify sensitivity in terms
of counting efficiency. To assess the inherent resolution,
examine at least three points on the counting efficiency
curve.
2) Make sure the manufacturer has not set the detector too
close to the noise level in order to stretch sensitivity
beyond allowable limits. To accomplish this, always accompany
the sensitivity/counting efficiency specification with a
minimum acceptable false count level.
3) Be aware of the degradation of sensitivity, resolution,
and accuracy when counting real world particles. These particles
appear in all shapes and refractive indices. Accordingly,
don't put all the emphasis on the PSL sphere sensitivity.
4) Specify that the counter must "see" 100% of
the particles passing through the counter at the 100% point
(and above) on the counting efficiency curve. Some available
counters, although extremely sensitive, "see"
only about 80% of the particles in the sampled aerosol at
the 100% point on the counting efficiency curve.
References
1) Martens, A.E., "Errors
in Measurements and Counting of Particles Using Light Scattering",
APCA Journal, Vol. 18, No. 10, October 1968.
2) Quenzel, H., "Influence of Refractive Index on the
Accuracy of Size Determination of Aerosol Particles with
Light-Scattering Aerosol Counters", Applied Optics,
Vol. No. 8, No.1, Ian, 1969.
3) Willeke, K, Liu, B.Y.H., "Single Particle Optical
Counter Principle and Application", University of Minnesota,
Particle Technology Lab Publication No. 264, May, 1975.
4) Burkman, D.C. et al, "Understanding and Specifying
the Sources and Effects of Surface Contamination in Semiconductor
Processing", Microcontamination, p. 57, November, 1988.
5) Ranade, M.B.,"Properties of Airborne Particles",
Prof. Dev. Course, University of Maryland, Mach 1989.
6) Whitby K.T, Vomela, R.A., "Response of Single Particle
Optical Counters to Nonideal Particles". Environmental
Science and Technology, Vol. 1, Oct, 1967.
7) Hoenig, S.A., "Where Are We and Where Do We Go Next?",
Microcontamination, p.22, April 1989.
8) Cooke, D.D., and M. Kerker, "Response Calculations
For Light Scattering Aerosol Particles", Applied Optics
Vol. 14 pp.734-739, March 1975.
9) Procedural Standards for Certified Testing of Cleanrooms,"
National Environmental Balancing Bureau, October, 1988.
10) "Federal Standard, Cleanroom and Workstation Requirements,
Controlled Environment," (FEDSTD-2091), June 15, 1988.
11) Hodkinson, J.R., "The Optical Measurement of Aerosols",
Aerosol Science, ch. Xi Academic Press, 1966.